Název: Methodology of camera spectral response measurement
Autoři: Melich, Radek ; Václavík, Jan ; Žídek, Karel
Typ dokumentu: Výzkumné zprávy
Rok: 2016
Jazyk: eng
Abstrakt: Spectral response of the camera will be measured using monochromator in conjunction with an integration sphere. Camera without objective will be placed in one port of the integrating sphere and light beam from monochromator will be fed into perpendicular port. Sensor of the camera sensor will be therefore uniformly illuminated by diffused light from monochromator. Proposed FWHM of the monochromator is 15 nm as a trade-off between light intensity provided by monochromator and deformation of spectral characteristic (Picture XY). Diameter of the integrating sphere has to be larger than 100 mm to ensure uniform illumination over sensor area.
Klíčová slova: Camera spectral response

Instituce: Ústav fyziky plazmatu AV ČR (web)
Informace o dostupnosti dokumentu: Dokument je dostupný v příslušném ústavu Akademie věd ČR.
Původní záznam: http://hdl.handle.net/11104/0294183

Trvalý odkaz NUŠL: http://www.nusl.cz/ntk/nusl-393194


Záznam je zařazen do těchto sbírek:
Věda a výzkum > AV ČR > Ústav fyziky plazmatu
Zprávy > Výzkumné zprávy
 Záznam vytvořen dne 2019-03-14, naposledy upraven 2021-11-24.


Není přiložen dokument
  • Exportovat ve formátu DC, NUŠL, RIS
  • Sdílet